Datacolor Spectrocolorimeter
Model #1005 |
Gretag Macbeth |
X-Rite |
| |
/  |
Spectrolino® |
Eye-One® Pro |
DTP41 Series II Autoscan |
Pulse™ |
| Price |
$599 / $499 |
$1,495 |
$1,400 |
$2,495 |
Appx. $1,000 |
| Measuring Geometry |
45º/0º - Dual Channel |
45º/0º, DIN 5033 |
45º/0º, DIN 5033 |
Reflection 45°/0° per ANSI/ISO 5-4 (IT2.17) |
0°/45°per ANSI / ISO 5.4 |
| Light Source |
LED array |
Tungsten, Type A |
Tungsten, Type A |
Gas Pressure @ 2850°K |
Gas pressure tungsten @ 2850°K |
| Measurement Aperture |
7 mm diameter illuminated; 4 mm diameter measured |
4 mm |
4.5 mm |
1.8mm (.070 in) in scan direction x 2.5mm (.097 in) wide |
3.2 mm |
| Warm-Up Time |
None |
Not Specified |
Not Specified |
None |
Not Specified |
| Measurement Time |
<1 second |
1 - 2.5 second |
Not Specified |
Approx. 0.25 sec/patch (7mm patch) |
< 2 seconds per 30 patch strip with Pathfinder, Integrated velocity feedback control system |
| Short-term Repeatability |
DE*94 <= 0.01 with respect to the mean CIELab value of 10 measurements every second on white. |
0.03 ΔE*CIELab (D50, 2°), mean value of 10 measurements every 10 s on white |
DE*94 <= 0.1 (D50,2°), with respect to the mean
CIELab value of 10 measurements every 3 seconds on white |
0.2 ΔE max.; ±0.01 Dmax on white |
0.2 DE*94 max.;±0.01 Dmax. on white |
| Inter-Instrument Agreement |
Average 0.4 DE*94, 1.0 Max DE*94
Measured on 12 BCRA tiles at 22°C. |
Typically 0.3 ΔE*CIELAB (D50, 2°), average based on 12 BCRA tiles
maximally 0.8 ΔE*CIELAB (D50, 2°) on 12 BCRA tiles |
Average 0.4 DE*94, max. 1.0 DE*94
(Deviation from GretagMacbeth manufacturing standard at 23°C for single measurement mode on 12 BCRA tiles (D50,2°) |
0.3 ΔE cmc avg. typical
(avg. based on 12 BCRA tiles) |
<0.3 DE*94 average
<0.6 DE*94 maximum |
| Output Function |
X Y Z
CIE L*a*b* |
CIE-XYZ, CIE-xyY, CIE-L*a*b*, CIE-L*
C*h*(a*b*), CIE-L*u*v*,
CIE-L*C*h*(u*v*),
LABmg, LCHmg, HunterLab, RxRyRz |
Luminance (cd/m2)
Chromaticity coord. x,y (CIE 1931) |
|
|
| Illuminants/Observers |
D50 CIE 2º, 1931 standard observer
D65 CIE 10º, 1964 standard observer |
D50, D65, A, C, D30...D300, F1...F12
2° & 10° Standard Observers |
|
A, C, D50, D55, D65, D75, F2, F7, F11 & F12
CIE 2° & 10° Standard Observers |
|
| Communication Interface |
USB 2.0 Full Speed |
RS-232 |
USB 1.1 |
RS-232, USB |
USB |
| Power Requirements |
USB powered |
90 to 270 V AC, 47 to 63 Hz |
USB powered |
12v DC, Universal 100–240 VAC;
50–60 Hz adapter; 30W |
Internal NiMH battery
USB charged via powered USB port (rapid charge with optional accessory AC adapter) |
| Dimensions of Measuring Head |
Height: 55.4 mm
Width: 97.5 mm
Length: 149.6 mm
Weight: 150 g |
Height: 63 mm
Width: 65 mm
Length: 155 mm
Weight: 230 g |
Height: 67 mm
Width: 66 mm
Length: 151 mm
Weight: 185 g |
Height: 88 mm
Width: 184 mm
Length: 114 mm
Weight: 890 g |
Height: 61 mm
Width: 61 mm
Length: 132 mm
Weight: 258 g |
| Environmental Requirements |
Operating Temperature:
+5º- +40º C
Relative Humidity:
max 85%, noncondensing |
|
|
"+10° (50°F) to +40°C (104°F) operating
30% to 85% RH non condensing" |
"+10°C (50°F) to 35°C (104°F);
30% to 85% RH (noncondensing) |